Skip to main content

CVE-2018-18984: CWE-311 Missing Encryption of Sensitive Data in Medtronic CareLink 9790 Programmer

Medium
VulnerabilityCVE-2018-18984cvecve-2018-18984cwe-311
Published: Fri Dec 14 2018 (12/14/2018, 15:00:00 UTC)
Source: CVE
Vendor/Project: Medtronic
Product: CareLink 9790 Programmer

Description

Medtronic CareLink and Encore Programmers do not encrypt or do not sufficiently encrypt sensitive PII and PHI information while at rest .

AI-Powered Analysis

AILast updated: 07/08/2025, 09:26:09 UTC

Technical Analysis

CVE-2018-18984 is a vulnerability identified in the Medtronic CareLink 9790 Programmer, a medical device programmer used to manage implantable cardiac devices. The core issue is the absence or insufficiency of encryption for sensitive Personally Identifiable Information (PII) and Protected Health Information (PHI) stored at rest on the device. This vulnerability is classified under CWE-311, which pertains to missing encryption of sensitive data. The lack of encryption means that if an attacker gains physical or logical access to the device's storage, they could potentially extract sensitive patient data without needing to bypass additional cryptographic protections. According to the CVSS v3.1 vector (CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N), the attack vector is physical access (AV:P), with low attack complexity (AC:L), no privileges required (PR:N), no user interaction (UI:N), unchanged scope (S:U), high impact on confidentiality (C:H), and no impact on integrity (I:N) or availability (A:N). This indicates that while the vulnerability does not affect the device's operation or data integrity, it poses a significant risk to patient confidentiality. The vulnerability affects all versions of the CareLink 9790 Programmer and was published in December 2018. There are no known exploits in the wild, and no patches have been linked or indicated, suggesting that remediation may require vendor intervention or device replacement. The vulnerability is particularly critical in healthcare environments where patient data privacy is strictly regulated and protected.

Potential Impact

For European organizations, this vulnerability poses a substantial risk to patient privacy and compliance with stringent data protection regulations such as the EU General Data Protection Regulation (GDPR). Unauthorized disclosure of PII and PHI could lead to severe legal penalties, reputational damage, and loss of patient trust. Healthcare providers using the CareLink 9790 Programmer may face challenges in safeguarding sensitive data, especially if devices are lost, stolen, or accessed by unauthorized personnel. The confidentiality breach could also impact clinical decision-making if patient data integrity is questioned, although this vulnerability does not directly affect data integrity or device availability. Additionally, the healthcare sector in Europe is a high-value target for cyber espionage and ransomware attacks, and unencrypted sensitive data could be leveraged in broader attack campaigns. The vulnerability's requirement for physical access somewhat limits remote exploitation but does not eliminate risk, particularly in environments with less stringent physical security controls or during device maintenance and transport.

Mitigation Recommendations

Given the absence of vendor patches, European healthcare organizations should implement compensating controls to mitigate this vulnerability. These include enforcing strict physical security measures around devices, such as secure storage, access logging, and controlled access to authorized personnel only. Encrypting backups and any extracted data from the device is essential. Organizations should conduct regular audits to ensure devices are accounted for and not tampered with. Network segmentation can limit exposure if the device interfaces with hospital networks. Additionally, organizations should engage with Medtronic to seek firmware updates or device replacements that address encryption deficiencies. Training staff on the importance of device security and data privacy is critical to prevent accidental exposure. Finally, organizations should review and update their incident response plans to include scenarios involving potential data breaches from medical devices lacking encryption.

Need more detailed analysis?Get Pro

Technical Details

Data Version
5.1
Assigner Short Name
icscert
Date Reserved
2018-11-06T00:00:00
Cisa Enriched
false
Cvss Version
3.1
State
PUBLISHED

Threat ID: 682f56360acd01a249263f62

Added to database: 5/22/2025, 4:52:06 PM

Last enriched: 7/8/2025, 9:26:09 AM

Last updated: 7/31/2025, 5:12:47 AM

Views: 21