CVE-2026-18844: Pulsetto Vagus Nerve Stimulator · CWE-912

Affects Pulsetto Vagus Nerve StimulatorSource CVE Database V5

0
Compare

Not known to be exploited · CVSS 7.2 High.

Data as of 9 Nov 2026, 23:24 UTC

CVSS 4.0 base score
7.2High
How severe the flaw is on the 0–10 CVSS scale.
EPSS
0.2% chance of exploitation in 30 days – higher than 10% of CVEsExploit Prediction Scoring System: the likelihood attackers will use this flaw soon.
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:N/SA:N

Exploitation evidence

  1. Disclosed: evidence found.

    CVE published 11 Aug 2026

  2. Detectable: no evidence on record.

    No public nuclei template indexed

  3. Weaponized: no evidence on record.

    No public exploit code indexed in Metasploit or Exploit-DB

  4. Exploited in the wild: no evidence on record.

    Not in CISA KEV and no source reports exploitation

  5. Observed by OffSeq Mirage: no evidence on record.

    No attack traffic recorded by OffSeq Mirage sensors

A missing step is not evidence of safety. Sources: cve.org (opens in a new tab), the Metasploit, Exploit-DB and nuclei-templates index as of 9 Nov 2026.

What to do

Affected software

ProductAffectedFixed in
PulsettoVagus Nerve StimulatorAffected
  • =All
Fixed in–

AI-suggested

Until a patch or official mitigation is available, restricting physical and Bluetooth access to the device is recommended to reduce risk.

Description

The firmware of the Pulsetto Vagus Nerve Stimulator accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on.