CVE-2026-18844: Pulsetto Vagus Nerve Stimulator · CWE-912
Affects Pulsetto Vagus Nerve StimulatorSource CVE Database V5
Not known to be exploited · CVSS 7.2 High.
Data as of 9 Nov 2026, 23:24 UTC
- EPSS
- 0.2% chance of exploitation in 30 days – higher than 10% of CVEsExploit Prediction Scoring System: the likelihood attackers will use this flaw soon.
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:N/SA:NExploitation evidence
Disclosed: evidence found.
CVE published 11 Aug 2026
Detectable: no evidence on record.
No public nuclei template indexed
Weaponized: no evidence on record.
No public exploit code indexed in Metasploit or Exploit-DB
Exploited in the wildExploited: no evidence on record.
Not in CISA KEV and no source reports exploitation
Observed by OffSeq MirageSeen by Mirage: no evidence on record.
No attack traffic recorded by OffSeq Mirage sensors
A missing step is not evidence of safety. Sources: cve.org (opens in a new tab), the Metasploit, Exploit-DB and nuclei-templates index as of 9 Nov 2026.
What to do
Affected software
| Product | Affected | Fixed in |
|---|---|---|
| PulsettoVagus Nerve Stimulator | Affected
| Fixed in– |
AI-suggested
Until a patch or official mitigation is available, restricting physical and Bluetooth access to the device is recommended to reduce risk.
Description
The firmware of the Pulsetto Vagus Nerve Stimulator accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on.