CVE-2026-7251: CWE-259 Use of hard-coded password in Eppendorf BioFlo 320
Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
AI Analysis
Technical Summary
CVE-2026-7251 identifies a critical security flaw in the Eppendorf BioFlo 320 where the VNC server uses a hard-coded password (CWE-259). This vulnerability enables unauthenticated remote attackers to access and control the device's user interface if remote access is enabled and the attacker knows the device's network address. The VNC communication is not encrypted, which could facilitate interception or replay attacks. The vulnerability affects all versions of the BioFlo 320. No vendor-provided patch or official remediation guidance is currently available.
Potential Impact
An attacker exploiting this vulnerability can gain full control over the BioFlo 320 device remotely without any user interaction or privileges. This includes access to all control panel features, potentially disrupting device operation or causing unauthorized changes. The lack of encryption in VNC traffic further exposes the device to interception risks. The CVSS 4.0 score of 9.3 reflects the critical severity and ease of exploitation over a network.
Mitigation Recommendations
Patch status is not yet confirmed — check the vendor advisory for current remediation guidance. Until an official fix is released, organizations should consider disabling remote VNC access if possible or isolating the device on a secure network segment to limit exposure. Monitoring network access to the device and restricting access to trusted hosts may reduce risk. Avoid exposing the BioFlo 320 to untrusted networks.
CVE-2026-7251: CWE-259 Use of hard-coded password in Eppendorf BioFlo 320
Description
Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
CVSS v4.0
Score 9.3critical
Weaknesses
AI-Powered Analysis
Machine-generated threat intelligence
Technical Analysis
CVE-2026-7251 identifies a critical security flaw in the Eppendorf BioFlo 320 where the VNC server uses a hard-coded password (CWE-259). This vulnerability enables unauthenticated remote attackers to access and control the device's user interface if remote access is enabled and the attacker knows the device's network address. The VNC communication is not encrypted, which could facilitate interception or replay attacks. The vulnerability affects all versions of the BioFlo 320. No vendor-provided patch or official remediation guidance is currently available.
Potential Impact
An attacker exploiting this vulnerability can gain full control over the BioFlo 320 device remotely without any user interaction or privileges. This includes access to all control panel features, potentially disrupting device operation or causing unauthorized changes. The lack of encryption in VNC traffic further exposes the device to interception risks. The CVSS 4.0 score of 9.3 reflects the critical severity and ease of exploitation over a network.
Mitigation Recommendations
Patch status is not yet confirmed — check the vendor advisory for current remediation guidance. Until an official fix is released, organizations should consider disabling remote VNC access if possible or isolating the device on a secure network segment to limit exposure. Monitoring network access to the device and restricting access to trusted hosts may reduce risk. Avoid exposing the BioFlo 320 to untrusted networks.
Technical Details
- Data Version
- 5.2
- Assigner Short Name
- icscert
- Date Reserved
- 2026-04-27T18:37:39.380Z
- Cvss Version
- 4.0
- State
- PUBLISHED
- Remediation Level
- null
Threat ID: 6a15e03a891d628fdc67d947
Added to database: 5/26/2026, 6:02:34 PM
Last enriched: 5/26/2026, 6:18:47 PM
Last updated: 5/26/2026, 10:51:27 PM
Views: 6
Community Reviews
0 reviewsCrowdsource mitigation strategies, share intel context, and vote on the most helpful responses. Sign in to add your voice and help keep defenders ahead.
Want to contribute mitigation steps or threat intel context? Sign in or create an account to join the community discussion.
Actions
Updates to AI analysis require Pro Console access. Upgrade inside Console → Billing.
Need more coverage?
Upgrade to Pro Console for AI refresh and higher limits.
For incident response and remediation, OffSeq services can help resolve threats faster.
Latest Threats
Check if your credentials are on the dark web
Instant breach scanning across billions of leaked records. Free tier available.