CVE-2025-54304: An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices
Source GCVE Database
Not known to be exploited · CVSS 9.8 Critical · end of life, no fix coming : replace or isolate affected systems; no fix will be released.
Data as of 25 Nov 2026, 04:05 UTC
- Patch
- None coming – the product is end of life
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HExploitation evidence
Disclosed: evidence found.
CVE published 4 Dec 2025
Detectable: no evidence on record.
No public nuclei template indexed
Weaponized: no evidence on record.
No public exploit code indexed in Metasploit or Exploit-DB
Exploited in the wildExploited: no evidence on record.
Not in CISA KEV and no source reports exploitation
Observed by OffSeq MirageSeen by Mirage: no evidence on record.
No attack traffic recorded by OffSeq Mirage sensors
A missing step is not evidence of safety. Sources: cve.org (opens in a new tab), the Metasploit, Exploit-DB and nuclei-templates index as of 24 Nov 2026.
What to do
No fix will be released: the affected product is end of life. Replace it, or isolate it from untrusted networks until you can.
AI-suggested
Mitigation options include isolating the device on a trusted network segment, disabling the X11 display server if possible, or preventing network access to port 6000. Network administrators should ensure that the device is not exposed to untrusted networks to reduce risk.
Description
Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices run an X11 display server on startup that listens on all network interfaces via port 6000. The default access control allows connections only from localhost and a specific IP (192.168.2.15). However, if the device obtains a different IP via DHCP, the display server becomes accessible to other network devices. This exposure allows attackers to gain root privileges and execute code remotely by interacting with the matchbox-desktop environment. This vulnerability affects only devices no longer supported by the vendor.
References
- nvd.nist.gov/vuln/detail/CVE-2025-54304 (opens in a new tab)
- assets.thermofisher.com/TFS-Assets/LSG/manuals/MAN0014388_IonOneTouch2Sys_UG.pdf (opens in a new tab)
- documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/Ion_OneTouch_2_and_Torrent_Suite_Software.pdf (opens in a new tab)
- thermofisher.com/order/catalog/product/4474779 (opens in a new tab)