Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices run an X11 display server on startup that listens on all network interfaces via port 6000. The default access control allows connections only from localhost and a specific IP (192.168.2.15). However, if the device obtains a different IP via DHCP, the display server becomes accessible to other network devices. This exposure allows attackers to gain root privileges and execute code remotely by interacting with the matchbox-desktop environment. This vulnerability affects only devices no longer supported by the vendor.