CVE-2025-15653: CWE-668 Exposure of Resource to Wrong Sphere in Dräger Zeus IE
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
AI Analysis
Technical Summary
CVE-2025-15653 is a local security vulnerability in Dräger Zeus IE and Zeus RS C500 anesthesia workstations. It involves exposure of device resources to an incorrect security sphere due to unprotected USB interfaces. Attackers with physical access can exploit this to compromise software integrity, impair therapy functions, manipulate processed data, or pivot to network-based attacks when the device is network-connected or linked to Dräger Service Connect. The vulnerability is classified under CWE-668 (Exposure of Resource to Wrong Sphere). No vendor advisory or patch information is available at this time.
Potential Impact
The vulnerability allows unauthorized individuals with physical access to compromise the integrity of critical medical device software. This can result in impaired therapy functions and manipulated device data, potentially affecting patient care. Additionally, the device could be leveraged as a pivot point for broader network attacks if connected to a network or Dräger Service Connect, increasing the risk of wider compromise.
Mitigation Recommendations
Patch status is not yet confirmed — check the vendor advisory for current remediation guidance. Until an official fix is available, restrict physical access to the affected devices and monitor for unauthorized USB interface use. Follow any future vendor instructions for updates or mitigations.
CVE-2025-15653: CWE-668 Exposure of Resource to Wrong Sphere in Dräger Zeus IE
Description
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
CVSS v4.0
Score 7.0high
Weaknesses
AI-Powered Analysis
Machine-generated threat intelligence
Technical Analysis
CVE-2025-15653 is a local security vulnerability in Dräger Zeus IE and Zeus RS C500 anesthesia workstations. It involves exposure of device resources to an incorrect security sphere due to unprotected USB interfaces. Attackers with physical access can exploit this to compromise software integrity, impair therapy functions, manipulate processed data, or pivot to network-based attacks when the device is network-connected or linked to Dräger Service Connect. The vulnerability is classified under CWE-668 (Exposure of Resource to Wrong Sphere). No vendor advisory or patch information is available at this time.
Potential Impact
The vulnerability allows unauthorized individuals with physical access to compromise the integrity of critical medical device software. This can result in impaired therapy functions and manipulated device data, potentially affecting patient care. Additionally, the device could be leveraged as a pivot point for broader network attacks if connected to a network or Dräger Service Connect, increasing the risk of wider compromise.
Mitigation Recommendations
Patch status is not yet confirmed — check the vendor advisory for current remediation guidance. Until an official fix is available, restrict physical access to the affected devices and monitor for unauthorized USB interface use. Follow any future vendor instructions for updates or mitigations.
Technical Details
- Data Version
- 5.2
- Assigner Short Name
- VulnCheck
- Date Reserved
- 2026-06-02T21:25:08.662Z
- Cvss Version
- 4.0
- State
- PUBLISHED
- Remediation Level
- null
Threat ID: 6a1f4fb5e29bf47b5009beb4
Added to database: 06/02/2026, 21:48:37 UTC
Last enriched: 06/10/2026, 11:03:05 UTC
Last updated: 07/31/2026, 19:22:53 UTC
Views: 88
Community Reviews
0 reviewsCrowdsource mitigation strategies, share intel context, and vote on the most helpful responses. Sign in to add your voice and help keep defenders ahead.
Want to contribute mitigation steps or threat intel context? Sign in or create an account to join the community discussion.
Actions
Updates to AI analysis require Pro Console access. Upgrade inside Console → Billing.
Need more coverage?
Upgrade to Pro Console for AI refresh and higher limits.
For incident response and remediation, OffSeq services can help resolve threats faster.
Latest Threats
Check if your credentials are on the dark web
Instant breach scanning across billions of leaked records. Free tier available.